personnel protection Rad B&P

Is this your test? Login to manage it. If not, you can generate a quiz just like it.

This is a non-interactive preview of the quiz content.

1.
1 point
In most fluoroscopic units the x-ray tube is located:
2.
1 point
Any wall in the exposure room that must stop primary radiation is considered a:
3.
1 point
Regulations require the source to skin distance not be less than ______ on mobile fluoroscopes
4.
1 point
All x-rays other than those in the primary beam
should be blocked before they can escape the tube housing. This blocking action is accomplished by lining the inside of the tube housing with _____
5.
1 point
The distance from the radiation source point to the image receptor is the:
6.
1 point
An on-off switch that requires continuous pressure to keep the switch ‘on’ is called a ______________ and is required for an exposure switch on fluoro machines
7.
1 point
Regulations require that the total filtration in the fluoroscopic beam be at:
8.
1 point
Regulations require the source to skin distance (SSD) not be less than ______ on fixed fluoroscopes
9.
1 point
Regulations require that the reproducibility for a set of exposure factors may not exceed a:
10.
1 point
Regulations require that the image intensifier be coupled with the x-ray tube and electrically interlocked so that the Fluoro x-ray tube cannot be energized when the image intensifier is in the “park position”
11.
1 point
Regulations require that linearity not exceed ________ change in dose between mA stations.
12.
1 point
Regulations require that a thyroid shield have a minimum lead equivalence of:
13.
1 point
Each x-ray tube should be provided with what mechanisms to ensure proper alignment of the beam with the image receptor?
14.
1 point
Regulation requires that the fluoro beam intensity as measured at the table top shall not exceed:
15.
1 point
A "Grave Danger Very High Radiation" sign is posted at entrance of a room where dose equivalent is in excess of:
16.
1 point
Regardless of what device is used to indicate SID, regulations require that it be accurate to:
17.
1 point
Protective gloves should be worn whenever the hands are placed into (or very near) the primary beam, they must contain a lead equivalent of:
18.
1 point
A "caution radiation area" sign is posted at entrance of a room where dose equivalent is in excess of:
19.
1 point
What are the factors that determine barrier design?
20.
1 point
Regulations require that the image intensifier cover have a:
21.
1 point
It is acceptable to not include an SID indicator on the unit to indicate exposure distance as long as there is a measuring device readily available in the room.
22.
1 point
Fluoro dose to personnel is also reduced by scheduling worker rotations.
23.
1 point
Regulations require that a fluoroscopic unit have a cumulative timer that will automatically stop the exposure after:
24.
1 point
Regulations require the bucky slot cover have a minimum lead equivalent of:
25.
1 point
Federal Regulations require that the diagnostic tube housing block leakage to less than ____________________from the housing.
26.
1 point
A sign posted at the entrance of a room where radioactive materials are used and/or stored.
27.
1 point
The portion of the non-useful beam that manages to escape the tube housing is called _______________
28.
1 point
Types of radiation relative to barrier design include:
29.
1 point
The ability of a machine to produce the same dose using the same mAs and kVp with different mA and time combinations.
30.
1 point
The control panel of the machine must indicate the conditions of the exposure including:
31.
1 point
All collimators must be periodically calibrated so that its shutters open to produce a light beam that superimposes the x-ray beam to:
32.
1 point
The opening along the length of the table between the table top and the table base to permit movement of the bucky tray:
33.
1 point
An imaging modality that uses x-rays and produces images in real time and used when evaluating the dynamics (movement) of anatomical structures.
34.
1 point
An audible and visible device must be present on the control to indicate to the operator when x-rays are being produced
35.
1 point
Leaded eyewear is also available and must have a minimum leaded equivalent of:
36.
1 point
An area with limited access for the purpose of protecting persons against undue risk from radiation exposure
37.
1 point
Any wall in the exposure room that must stop secondary or scatter radiation is considered to be a
38.
1 point
The ability of an x-ray machine to duplicate an exposure dose for any technique that is selected.
39.
1 point
Regulations from NCRP report 102 says leaded aprons have a minimum lead equivalent of:
40.
1 point
Regulations require that the protective curtain have a minimum lead equivalent of:
41.
1 point
A "Caution High Radiation" sign is posted at entrance of a room where dose equivalent is in excess of:
42.
1 point
Primary radiation shielding requires a __________ than secondary shielding.
43.
1 point
Any barrier that can stop the primary beam:
44.
1 point
When the exposure is made x-rays immediately move away from their source point in all directions. The x-rays are moving _________
45.
1 point
The x-ray tube housing is mounted on a support column called a
46.
1 point
Even though regulations required less, protective gloves are most commonly made with a:
47.
1 point
The design in which the collimator automatically limits the primary beam to match the size of the image receptor is _____________
48.
1 point
During fluoroscopy the primary protective barrier for the machine operator is the ________________
49.
1 point
__________ is only those x-rays that are moving in the direction of the patient that are of any value to the imaging process
50.
1 point
Because of the length of the fluoro exposure time the beam intensity must be _______ when compared to standard radiography beam intensity